Type |
Maximum thermometer |
Measurement |
90.0 F to 109.9 F / 32 C to 43.9 C |
Range |
Temp <90.0 F / 32 C : display <<L>> for low (too low) Temp> 109.9 F / 43.9 C: display <<H>> for high (too high) |
Measurement |
+/- .02 F/.01 C between 96.0 F to 107.6 F / 34 C to 42 C |
Accuracy |
An ambient temperature of 64.4 F to 82.4 F / 18 C to 28 C |
Self-Test |
Automatic internal check at a test value of 98.6 F / 37.7 C; if there is a deviation of >0.2 F /0.1 C “ERR” (error) |
Display |
Liquid crystal display (LCD) with three digits Smallest unit of display: 0.1 F/0.5C |
Signaling Tone |
For Signaling that the thermometer is ready to use and that the temperature increase is less than 0.2 F (0.1 C) / 16 seconds |
Memory |
For storing the last measured value |
Storage Temp. |
-50 F to 140 F |
Battery |
1.5/1.55V |
Reference to |
PREN12470-3:1997. clinical thermometers |
Standards |
ATSM standard |